| 仪器特点 1、微机全自动触控量热仪采用**计算机控制程序,结构紧凑,可快速无噪声地打印数据。可自动标定量热系统的热容量(能当量),测量发热量。输入硫、水分、氢等数据,即可换算并打印出弹筒发热量、高位发热量,低位发热量等结果。2、恒温式量热仪内筒采用片状桨叶的电动搅拌,外筒采用循环泵搅拌,使水温更均匀;增加了注水口、抽水口和自动调水温部分,使操作更简单更方便。仪器即可使用熔断式也可使用非熔断方式点火,使用方便自如。3、系统操作全过程汉字提示,按提示操作即可完成试验。
 试验室条件 试验室应设在一单独房间,不得在同一房间内同时进行其它试验项目。室温应尽量保持恒定,每次测定室温变化不应超过1℃,通常室温尽量保持在15~30℃范围为宜。室内应无强烈的空气对流,不应有强烈的热源和风扇等,试验过程中应避免开启门窗。试验室*好阳光照射,将量热仪放在不受阳光直射的地方。 Laboratory conditions The laboratory should be set up in a separate room,No other test items shall be carried out in the same room at the same time。Room temperature should be kept as constant as possible, and the change of room temperature should not exceed 1 ° C,It is generally advisable to keep the room temperature in the range of 15 ~ 30 °C. There should be no strong convection in the room,Should not have the intense heat source and the fan and so on, should avoid in the test process opens the door window。In the laboratory, the calorimeter is placed in a place where it is not exposed to direct sunlight。 突发故障 一、点火失败 1、*先检查保险管,保险的位置通常在机器的后面。 2、检查氧弹底下的三个支柱,顶盖上的点火头是否生锈,如生锈会造成接触不良,应用砂纸打磨去锈。 3、检查氧弹内点火丝是否卡紧,棉线是否连接上,氧气是否充足,如没有都会造成点火失败。 二、氧弹漏气 1、主要是检查密封圈的损坏,机器上一共有四种密封圈。 2、化验人员应向调试安装人员学习四张密封圈的拆换和位置所在。 日常维护 1、*先要保护环境的卫生。 2、要把机器盖打开,保持机器的干燥。 3、在温度低时,要保持机器内的水不能上冻。 Daily Maintenance 1、*First of all, we must protect the health of the environment。 2、To open the machine cover, keep the machine dry。 3、Keep the water in the machine from freezing when the temperature is low。  
标签:微机全自动触控量热仪,自动测温仪,检测煤炭大卡设备 |